Zefon Escort ELF Personal Air Sampling Pump
Zefon Escort ELF Personal Air Sampling Pump
Escort ELF air sampling pumps feature a state of the art electronic laminar flow sensor to provide a constant flow control with ± 2.5% regulation of flow rate (from 1-3 lpm) with automatic compensation for changes in battery voltage, temperature, altitude, and sample load. An internal secondary standard calibrates the pump continuously and needs to checked against a primary standard only once per month (or every 200 hours for coal mine dust sampling).
Features:
- Patented electronic laminar flow control.
- Internal secondary standard calibrates the pump continuously, maintaining OSHA compliance.
- Check against a primary standard only once a month or every 200 hours for coal mine dust sampling.
- Inlet filter traps moisture and particular matter.
- Stainless steel-filled plastic case provides EMI/RFI protection.
- Rated IP65 for resistance to water and dust ingress.
- Exceptionally quiet operation.
- Flow-Fault Indicator shows any sample inlet blockage.
- Run/Hold feature allows “time out” during sampling process.
- Elapsed time readout and low-battery warning.
- Low flow and dual tube sampling available with Gemini Twin-Port Sampler.
- 2 Year Warranty.
Approvals
- NIOSH certified for coal mine dust sampling (TC-74-030)
- MSHA certified as intrinsically safe for underground use (Approval # 2G-3924-1).
- UL approved as intrinsically safe for use in hazardous locations, Class I, Groups A, B, C, D; Class II, Groups E, F and G; Class III, Division I locations.
- ATEX Approved models available. Contact Zefon or your Zefon distributor for details.
· Specifications
Operating Flow Control |
Volumetric flow rate held within ± 2.5% of set-point over the 1-to-3 lpm operating range (± 5% to 0.5 lpm), with automatic compensation for battery voltage, altitude, temperature and sample load changes. |
Flow Rate |
Flow rate adjustable between 0.5 and 3 lpm; lower flow rates attainable (to 1 mlpm) with optional Gemini Twin-Port Sampling accessory. |
Flow Indication |
Liquid Crystal Display provides 0.01 lpm resolution. |
Operating Range |
30 inches of water load up to 2 lpm and 10 inches up to 3 lpm. |
Flow Blockage Detection |
Flow-Fault LED comes on immediately when block is detected. Pump shuts down within 90 seconds if block is not cleared. |
Elapsed Time Readout |
To 999 minutes in increments of 1 minute. Holds the last reading after either fl ow blockage or low battery shutdown and when pump is off or in hold mode. |
Operating Time |
Varies with flow rate and sampling device loading. |
Operating Temperature Limits |
32°-113°F (0°-45°C) |
Physical Weight |
23 oz. (652 gm), with battery pack. |
Dimensions |
2.25” D x 4.25” H x 4” W |
Power Supply |
4.8-volt battery pack of four NiMH cells. (Units Mfg before Oct 2011 contain NiCd cells) |
Battery Pack Capacity |
2.8 amp hours (NiMH version), 2.4 amp hours (NiCD version) |
Battery Pack Recharge Time |
14-16 hours (overnight) with Omega™ Charger. |
Typical Battery Pack Life |
300 or more charging cycles |